Products Search
>
GC
>
GC-2030
>
Error/Warning
>
CARx prim is out of range[Error Code:2138 2139 2140 2212]
Back
No : 6017
Date of Issue : 2021/05/18 15:36
Last Updated : 2021/06/03 11:51
Print
CARx prim is out of range[Error Code:2138 2139 2140 2212]
The carrier gas supply pressure is incorrect.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Check that the carrier gas supply pressure is within the primary pressure range in the configuration described below.
Screen_Other-Configuration.pdf
An error will occur if the primary pressure goes below the following values.
[300-500 kPa] : 150 kPa
[500-900 kPa] : 350 kPa
[900 kPa-] : 750 kPa