Products Search
>
GC
>
GC-2030
>
Error/Warning
>
SCD out pyrotube use time warn.[Error Code:9074]
Back
No : 5954
Date of Issue : 2021/05/18 15:37
Last Updated : 2021/06/03 15:53
Print
SCD out pyrotube use time warn.[Error Code:9074]
The number of analysis exceeded the set threshold.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Replace the outer pyrotube and then reset the counter.
※Replace the outer pyrotube
MaintenanceHelp_SCD-Outer-Pyrotube-Replacement.pdf
※Reset the counter
[HOME] - [Detector] - [Submenu] - [SCD Parts Counter] to display [SCD Parts Counter] screen.
Screen_SCD-Parts-Counter.pdf