Products Search
>
GC
>
GC-2030
>
Error/Warning
>
SCD in pyrotube use time warn[Error Code:9073]
Back
No : 5953
Date of Issue : 2021/05/18 15:37
Last Updated : 2021/06/03 15:00
Print
SCD in pyrotube use time warn[Error Code:9073]
The number of analysis exceeded the set threshold.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Replace the inner pyrotube and then reset the counter.
※Replace the inner pyrotube
MaintenanceHelp_SCD-Inner-Pyrotube-Replacement.pdf
※ Reset the counter
[HOME] - [Detector] - [Submenu] - [SCD Parts Counter] to display [SCD Parts Counter] screen.
Screen_SCD-Parts-Counter.pdf