SCD vacuum pump use time warn. [Error Code:9071]
The number of analysis exceeded the set threshold.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Replace or maintain the vacuum pump and then reset the counter.
SCD vacuum pump replacement
MaintenanceHelp_SCD-Vacuum-Pump-Onspection.pdf
Counter reset
Press [HOME] > [Detector] > [Submenu] > [SCD Parts Counter],
and then display [SCD Parts Counter] screen and reset the counter.
Screen_SCD-Parts-Counter.pdf