• No : 5932
  • Date of Issue : 2021/05/18 15:36
  • Last Updated : 2021/06/08 11:22
  • Print

Clean up program time error [Error Code:4303]

Program time is longer than the upper limit "9999.99 min".
Category : 

Answer

Do not set the total time longer than 9999.99 min.
The analysis can be performed even after this message is displayed, however, the program finishes after 9999.99 min utes elapses.