Products Search
>
GC
>
GC-2030
>
Error/Warning
>
Gas Selector# purge set point not reached [Error Code:234...
Back
No : 5920
Date of Issue : 2021/05/18 15:37
Last Updated : 2021/06/12 01:28
Print
Gas Selector# purge set point not reached [Error Code:2348 2349]
The pressure does not increase to the set value due to an error in pressure control (leak).
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Check whether gas is supplied and whether gas is not leaking from connections. Gas leaks may be occurring in several locations (other than displayed site).