CARm AFC control irregularity [Error Code:2134 2135 2136 2137 2208 2209 2210 2211]
Each unit does not operate correctly.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Check that the supply pressure to the AFC is within the range below.
Check that the supply pressure is sufficient for the AFC gas to flow.
TechnicalDocument_AFCnAPC-Control-Ranges-TotalFlow-n-InletPressure.pdf
Check that the gas supply pressure is stable.
If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.