Products Search
>
GC
>
GC-2030
>
Error/Warning
>
SCD PSU-24V error [Error Code:6124]
Back
No : 5906
Date of Issue : 2021/05/18 15:37
Last Updated : 2021/06/12 00:54
Print
SCD PSU-24V error [Error Code:6124]
The power source may have failed.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.