Error/Warning
>
Room temperature is out of range [Error Code:5]
Back
No : 5901
Date of Issue : 2021/05/18 15:37
Last Updated : 2021/06/12 00:43
Print
Room temperature is out of range [Error Code:5]
If this error is displayed when the actual room temperature and air pressure is within the operation range, sensors installed on the instrument may have failed.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.