CARx calc. prss out of range [Error Code:5006 5007 5008 5070 5071 5072 5101 5105]
The column inlet pressure calculated from the inputted linear velocity and column flow is outside the set range.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Set the linear velocity and column flow so that the column inlet pressure is within the AFC control range.
TechnicalDocument_AFCnAPC-Control-Ranges-TotalFlow-n-InletPressure.pdf
Check the inputted column information (inner diameter, length, and film thickness).
Check the initial column temperature setting.