• No : 5875
  • Date of Issue : 2021/05/18 15:37
  • Last Updated : 2021/06/04 14:09
  • Print

DET#1 2 3 4 FTD adjustment failed [Error Code:4149 4150 4151 4152]

It cannot be adjusted because FTD power (%) is [0] or [100].
Category : 

Answer

Check that connection of electrode connector is correct and collector fixing knurled screws are tightened.

When connections are correct, recondition or replace FTD alkaline source.
 
If the error cannot be resolved after the above actions, contact your Shimadzu sales/service representative.