SCD furnace abs. press was exceeded upper limit [Error Co...
Back
No : 5856
Date of Issue : 2021/05/18 15:38
Last Updated : 2021/06/04 11:45
Print
SCD furnace abs. press was exceeded upper limit [Error Code:6112]
The pressure of the furnace is high.
The inner pyrotube may be clogged.
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Replace the inner pyrotube.
MaintenanceHelp_SCD-Inner-Pyrotube-Replacement.pdf
If the error is displayed even after the tube is replaced, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.