Products Search
>
GC
>
GC-2030
>
Error/Warning
>
SCD furnace temperature controller error [Error Code:6105]
Back
No : 5853
Date of Issue : 2021/05/18 15:38
Last Updated : 2021/06/04 11:40
Print
SCD furnace temperature controller error [Error Code:6105]
The temperature of the furnace does not increase even after [Heater] is set to [On].
Category :
Products Search
>
GC
>
GC-2030
>
Error/Warning
Answer
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.