The AA-7800/7000 series instruments are equipped with a safety device to prevent flashback, but flashback may still occur when using the flame method. If flashback occurs, stop using the instrument immediately and implement the measures in the event of an emergency.
WARNING: When flashback occurs, the burner head can be thrown 10 cm upward.
NOTE:
If the instrument detects flashback, ignition will become impossible.
Before using the instrument again, request
your Shimadzu representative to inspect it.
Once the representative has inspected the instrument and confirmed that there is no problem with it, he will restore it to the condition in which ignition is possible.