Turn on the power of the spectrophotometer, then initialization start and display results in the order of the check
items.
(Without display instrumets, after connect to software, initialization window appears to display results in the order of the check items.)
• When an error is detected in ROM/RAM or motor.
Turn off the instrument's power switch, and turn it on again to initialize the instrument.
• When an error is detected in relation to lamp energy or wavelength.
• Is there anything obstructing the beam in the sample compartment cell holder?
Turn off the instrument's power switch.
Remove the obstruction and turn on the power switch again.
• Is the sample compartment cover opened?
Turn off the instrument's power switch.
Close the sample compartment cover and turn on the power switch again to initialize the instrument.
• Is the light source lamp lit?
You can check it directly by checking the lamp sign on the display or opening the light source cover.
If the light source lamp is not lit, replace the light source lamp.
Click here for replacing the light souce.