Cause 1: Carrier gas flows?
Action 1: If carrier gas is not flowing, please let the carrier gas flow.
Cause 2: Carrier gas leaks.
Action 2: Tighten any leaking connections.
Click here for Checking for Leaks.
Cause 3: Proper column installation.
Action 3: Install the column correctly.
Cause 4: Temperature of column, injection port or interface has dropped.
Action 4: Increase the temperature of each heated zone to the appropriate value.
Cause 5: Appropriate column selection.
Action 5: Change to an appropriate column (check column polarity, etc.).
Cause 6: Split ratio too large.
Action 6: Decrease split ratio.
Cause 7: Sample concentration or injection amount too low.
Action 7: Increase sample concentration or inject a greater volume of sample.
Cause 8: Filament open (or burned out).
Action 8: Change or replace the filament.
Click here for Filament Replacement Procedure.
Cause 9: Detector voltage too low.
Action 9: Increase detector voltage gradually.
Cause 10: Appropriate analytical conditions.
Action 10: Correct the method parameters.
Cause 11: MS properly tuned.
Action 11: Perform autotuning or adjust the instrument.