• No : 4136
  • Date of Issue : 2020/12/02 16:46
  • Last Updated : 2020/12/07 11:07
  • Print

Error Code:[0x5AE6]

Error Message: PDA detector: D2 Lamp failed to light. Replace the D2 lamp.
Category : 

Answer

«Main Cause»
The D2 lamp failed to illuminate.
 
«Countermeasure»
Replace the D2 lamp.
Click here for Replacement method of Lamp (HPLC)