Products Search
>
HPLC
>
Detector
>
Error Code:[0x5AE4]
Back
No : 4134
Date of Issue : 2020/12/02 16:46
Print
Error Code:[0x5AE4]
Error Message: PDA detector: D2 Lamp PCB is not operating normally.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
The illumination circuit is not functioning properly.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.