Error Code:[0x5AD9]
Error Message: PDA detector: The thermosensor of monochromator did not operate normally, so cell temperature control was stopped.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
A broken circuit may have occurred in the monochromator temperature sensor.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.