Error Code:[0x5AD6]
Error Message: PDA detector: The lamp was turned off because the inside fan did not operate normally.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
A rotation problem was detected for the internal fan.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.