• No : 4120
  • Date of Issue : 2020/12/02 16:46
  • Print

Error Code:[0x5AC8]

Error Message: PDA detector: The lamp was turned off because the temperature of the lamp housing was too high.
Category : 

Answer

«Main Cause»
The lamp housing became hot.
 
«Countermeasure»
Implement corrective measures in the following sequence.
(1) Check whether the fan on the back of the instrument is blocked.
(2) Clean or replace the filter on the back of the instrument.
If that does not resolve the error, switch the instrument power OFF and contact a Shimadzu sales or service representative.