Error Code:[0x5A84]
Back
No : 4113
Date of Issue : 2020/12/02 16:46
Print
Error Code:[0x5A84]
Error Message: PDA detector: The calibration data for the Undefined slots in PDA detector cell cannot be read normally. Perform calibration again.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
The calibration information for Undefined slots is the default value.
«Countermeasure»
Register the calibration information using PDAUtility.