Error Code:[0x5A82]
Error Message: PDA detector: Lamp housing heater did not operate normally, so the cell temperature control was stopped.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
A broken circuit may have occurred in the lamp housing heater.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.