Error Code:[0x5A61]
Error Message: PDA detector: The temperature control for the cell was stopped because the temperature was over the setting value.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
The temperature-controlled cell control circuit is not functioning properly.
«Countermeasure»
Switch the instrument power OFF and contact a