Products Search
>
HPLC
>
Detector
>
Error Code:[0x5A45]
Back
No : 4093
Date of Issue : 2020/12/02 16:45
Print
Error Code:[0x5A45]
Error Message: PDA detector: Intensity is too high because the exposure time of PDAs is too long. Run recalibration on exposure time.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
Light level exceeded the threshold value due to the long accumulation time.
«Countermeasure»
Calibrate the accumulation time.