Error Code:[0x58CC] ERROR W LAMP NOT CONNECTED
Back
No : 4071
Date of Issue : 2020/12/02 16:45
Last Updated : 2020/12/07 11:12
Print
Error Code:[0x58CC] ERROR W LAMP NOT CONNECTED
Error Message: UV detector: The connection of W lamp failed to be checked. Check the connection of W lamp. If the error is not resolved, replace W lamp.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
The W lamp is not connected or the illumination circuit is not functioning properly.
«Countermeasure»
Implement corrective measures in the following sequence.
(1) Check the W lamp connection.
(2) Replace the W lamp.
Click here
for Replacement method of Lamp (HPLC)
If that does not resolve the error, switch the instrument power OFF and contact a
Shimadzu sales or service representative
.