Products Search
>
HPLC
>
Detector
>
Error Code:[0x5885] FATAL REAR FAN
Back
No : 4063
Date of Issue : 2020/12/02 16:45
Print
Error Code:[0x5885] FATAL REAR FAN
Error Message: UV detector: The lamp was turned off because the back fans did not operate normally.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
A rotation problem was detected for the fan on the back of the instrument.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.