Error Code:[0x5870] ERROR CELL TEMP SENSOR
Back
No : 4059
Date of Issue : 2020/12/02 16:45
Print
Error Code:[0x5870] ERROR CELL TEMP SENSOR
Error Message: UV detector: Temperature of the flow sensor did not operate normally, so cell temperature control was stopped.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
A broken circuit may have occurred in the cell temperature sensor.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.