Products Search
>
HPLC
>
Autosampler
>
Error Code:[0x5288] ERROR FCV7 ERROR
Back
No : 4017
Date of Issue : 2020/12/02 16:45
Print
Error Code:[0x5288] ERROR FCV7 ERROR
Error Message: Autosampler: The Rotary Valve 7 in the autosampler has detected an error. Check Rotary Valve 7.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
An error was detected from the FCV7 unit connected to the autosampler.
«Countermeasure»
Check the FCV7 unit. If that does not resolve the error, switch the power OFF and contact a
Shimadzu sales or service representative
.