Error Code:[0x5268] ERROR PFILE NO ENDIF
Error Message: Autosampler: There is no ENDIF command corresponding to the IF command of the pretreatment program. Add an ENDIF command.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The ENDIF command corresponding to an IF command in the pretreatment program was not found.
«Countermeasure»
Add an ENDIF command.