Products Search
>
HPLC
>
Autosampler
>
Error Code:[0x5265] ERROR PFILE ASSIGN RVAR
Back
No : 3982
Date of Issue : 2020/12/02 16:44
Print
Error Code:[0x5265] ERROR PFILE ASSIGN RVAR
Error Message: Autosampler: The value assigned to the variable of the pretreatment program exceeds the maximum value of 65536. Correct the pretreatment program.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The value substituted for a variable in a calculation command in the pretreatment program exceeded the maximum value 65536.
«Countermeasure»
Revise the pretreatment program so that assigned values do not exceed the maximum value.