Products Search
>
HPLC
>
Autosampler
>
Error Code:[0x5264] ERROR PFILE CALC OVER
Back
No : 3981
Date of Issue : 2020/12/02 16:44
Print
Error Code:[0x5264] ERROR PFILE CALC OVER
Error Message: Autosampler: The result of the calculation command in the pretreatment program exceeds 65536. Correct the pretreatment program.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The result of a calculation command in the pretreatment program exceeded the maximum value 65536.
«Countermeasure»
Revise the pretreatment program so that calculation results do not exceed the maximum value.