Products Search
>
HPLC
>
Autosampler
>
Error Code:[0x5259] ERROR PFILE CALL TIMES
Back
No : 3970
Date of Issue : 2020/12/02 16:44
Print
Error Code:[0x5259] ERROR PFILE CALL TIMES
Error Message: Autosampler: The call count of the CALL command in the pretreatment program is over 19. Make the count 19 or less.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The CALL command calls more than 19 times in the pretreatment program.
«Countermeasure»
Set it to 19 or fewer times.