Error Code:[0x523A] ERROR PFILE PURGE SPEED
Error Message: Autosampler: The aspiration speed of the PURGE.RP or PURGE.ML command in the pretreatment program is out of range. Change the aspiration speed to within the range.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The suction speed specified by the PURGE.RP or PURGE.ML command in the pretreatment program exceeds the setting range.
«Countermeasure»
Change the setting to a value within the specified range.