Products Search
>
HPLC
>
Autosampler
>
Error Code:[0x5239] ERROR PFILE PURGE VOL
Back
No : 3938
Date of Issue : 2020/12/02 16:44
Print
Error Code:[0x5239] ERROR PFILE PURGE VOL
Error Message: Autosampler: The liquid volume of the PURGE.RP or PURGE.ML command in the pretreatment program is out of range. Change the liquid volume to within the range.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
The liquid volume specified by the PURGE.RP or PURGE.ML command in the pretreatment program exceeds the setting range.
«Countermeasure»
Change the setting to a value within the specified range.