Products Search
>
HPLC
>
Solvent Delivery Module
>
Error Code:[0x9117] FATAL INSIDE FAN
Back
No : 3878
Date of Issue : 2020/12/02 16:44
Print
Error Code:[0x9117] FATAL INSIDE FAN
Error Message: Pump: The pumping was stopped because the motor-cooling fan detected an error.
Category :
Products Search
>
HPLC
>
Solvent Delivery Module
Answer
«Main Cause»
A motor cooling fan rotation problem was detected.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.