Products Search
>
HPLC
>
Solvent Delivery Module
>
Error Code:[0x910B] FATAL LEAK SENSOR
Back
No : 3868
Date of Issue : 2020/12/02 16:43
Print
Error Code:[0x910B] FATAL LEAK SENSOR
Error Message: Pump: The pumping was stopped because leak sensor cannot be recognized correctly.
Category :
Products Search
>
HPLC
>
Solvent Delivery Module
Answer
«Main Cause»
A broken circuit may have occurred in the leak sensor.
«Countermeasure»
Switch the instrument power OFF and contact a
Shimadzu sales or service representative
.