Error Code:[0x00C5] ERR BG2 TOO HIGH
Error Message: Ch2 background is too high.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
GAIN1/GAIN2 was set to 1 and the background level exceeded 100μScm-1 when autozero was performed. GAIN1/GAIN2 should be 2 or larger if background is more than 100μScm-1.
«Countermeasure»
Set GAIN1/GAIN2 to 2 or higher.