Error Code:[0x8B43] SFE ERR T.MAX TUBE
Error Message: The temperature of the preheater exceeded upper temperature limit, so the cell temperature control was turned stop.
Category :
Products Search
>
HPLC
>
Supercritical Fluid Chromatograph
Answer
«Main Cause»
The temperature of the pre-heater exceeded the [Maximum Temperature] of the instrument parameters.
«Countermeasure»
Maximum temperature must be specified 5ºC higher than the extraction vessel temperature.