• No : 3677
  • Date of Issue : 2020/12/02 16:31
  • Last Updated : 2020/12/07 11:22
  • Print

Error Code:[0x89D4]

Error Message: D2 lamp overvoltage.
Category : 

Answer

«Main Cause»
(1)D2 lamp malfunction.
(2)Malfunction inside the instrument.
 
«Countermeasure»
(1) Replace the D2 lamp.
    Click here for Replacement method of Lamp (HPLC)
(2) Turn the power OFF and contact a Shimadzu sales or service representative.