Products Search
>
HPLC
>
Detector
>
Error Code:[0x89C8]
Back
No : 3668
Date of Issue : 2020/12/02 16:31
Print
Error Code:[0x89C8]
Error Message: The light source room is overheated.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
A probable cause is that the cooling fan blast flow is poor, or the cooling fan or heater circuit malfunctioned.
«Countermeasure»
Ensure at least 15 mm of space on the sides of the PDA detector. If this does not remedy the problem, contact a
Shimadzu sales or service representative
.