Products Search
>
HPLC
>
Detector
>
Error Code:[0x8962]
Back
No : 3655
Date of Issue : 2020/12/02 16:31
Print
Error Code:[0x8962]
Error Message: The cell current is excessive.
Category :
Products Search
>
HPLC
>
Detector
Answer
«Main Cause»
Temperature-controlled flow cell malfunctioned, control circuit malfunctioned.
«Countermeasure»
Turn the power OFF and contact a
Shimadzu sales or service representative
.