• No : 3637
  • Date of Issue : 2020/12/02 16:30
  • Print

Error Code:[0x0030] ERROR P.MAX

Error Message: Solvent delivery unit :pump's pressure has exceeded P.Max. Check for flow line clogging and P.Max Settings.
Category : 

Answer

«Main Cause»
The discharge pressure of pump exceeds the Settings [P.Max]. Pumping will automatically stop. However, if "S-PROT" of auxiliary operation is Settings, set flow rate to 1/2 for pumping.
 
«Countermeasure»
Inspection the flow line clog. If there is no \n abnormality, Settings the appropriate values in [P.Max].