Error Code:[0x8547] STOP COOLER M
Error Message: The rack has not been set for more than 30 minutes in the rack plate M, so the rack plate M cooler was turned off. Set rack to restart the sample cooler cell temperature control.
Category :
Products Search
>
HPLC
>
Autosampler
Answer
«Main Cause»
Since sample cooler M has operated without sample rack for more than 30 minutes, sample cooler M is stopped.
«Countermeasure»
Insert a sample rack into rack plate M. Or, stop sample cooler M.