• No : 3421
  • Date of Issue : 2020/12/02 16:10
  • Last Updated : 2020/12/07 11:16
  • Print

Error Code:[0x93C7]

Error Message: The D2 lamp cannot light because the heater current of the detector is too large.
Category : 

Answer

«Main Cause»
There is a fault in the D2 lamp or a fault in the lighting circuit.
 
«Countermeasure»
Replace the D2 lamp.
Click here for Replacement method of Lamp (HPLC)
If the error is not resolved, press the (power switch) to turn the power OFF and contact a Shimadzu sales or service representative.