(SMX-1010/1020) There is unevenness or noise in X-ray images

(SMX-1010/1020) There is unevenness or noise in X-ray images
Category : 

Answer

Cause: Unevenness and noise in fluoroscopic images may be caused by deterioration of the flat panel detector over time.
Action: Calibrate the flat panel detector.
  Click here for "(SMX-1010/1020) Performing Flat Panel Detector Calibration".
  If the condition does not improve, contact a Shimadzu sales or service representative.