(PPSQ) Replacing O-rings

(PPSQ) Replacing O-rings
Category : 

Answer

O-rings are used to seal reagent/solvent bottles and the waste bottle.
If the O-ring in a reagent/solvent bottle is cracked or cut, the pressure may decrease and it may not be possible to deliver the reagent/solvent.
If the O-ring in the waste bottle is cracked or cut, a bad odor may be released.
Occasionally, when replacing a reagent/solvent or when processing waste, inspect the
O-ring damage such as cuts and cracks. If the O-ring is damaged, replace it immediately.
Part Number: 290-36002-06  O-ring, R1
Part Number: 290-36002-08  O-ring,R2,S2,S3,S4B
Part Number: 036-15672-06  O-ring,R3
Part Number: 036-15720-03  O-ring,R4
Part Number: 036-15711-06  O-ring, Wast