(MCE) (ERROR)Undercurrent error
(MCE) (ERROR)Undercurrent error
- Category :
Answer
<Details>
An insufficient current value has been detected.
An insufficient current value has been detected.
<Typical Causes>
• The microchip was clogged.
• Air bubbles were mixed into the separation channel.
• Poor contact of electrode contacts on the tip cover.
<Instrument Operations>
Analysis is aborted for the sample being analyzed at the time the error occurred. Subsequent sample analyses proceed, ending with automatic rinsing of the microchips.
Analysis is aborted for the sample being analyzed at the time the error occurred. Subsequent sample analyses proceed, ending with automatic rinsing of the microchips.
<Corrective Measures>
• Return the drives to home position as in "Moving All Axes to Home Position".
Click here for "(MCE) Moving All Axes to Home Position".
• Open the top cover and chip cover and remove the affected microchip.
• Rinse the microchips according to the procedure described in "Inspecting and Washing Microchip Reservoirs".
• Open the top cover and chip cover and remove the affected microchip.
• Rinse the microchips according to the procedure described in "Inspecting and Washing Microchip Reservoirs".
Click here for "(MCE) Inspecting and Washing Microchip Reservoirs (manually)".
• Perform maintenance on the chip cover.
Click here for "(MCE) Maintenance of Chip Cover".