• Font Size
  • S
  • M
  • L
  • No : 8382
  • Date of Issue : 2022/02/18 10:23
  • Last Updated : 2022/02/25 15:57
  • Print

(MCE) (ERROR)Undercurrent error

(MCE) (ERROR)Undercurrent error
Category : 

Answer

<Details>
An insufficient current value has been detected.

<Typical Causes>
• The microchip was clogged.
• Air bubbles were mixed into the separation channel.
• Poor contact of electrode contacts on the tip cover.
 
<Instrument Operations>
Analysis is aborted for the sample being analyzed at the time the error occurred. Subsequent sample analyses proceed, ending with automatic rinsing of the microchips.

<Corrective Measures>
• Return the drives to home position as in "Moving All Axes to Home Position".
  Click here for "(MCE) Moving All Axes to Home Position".
• Open the top cover and chip cover and remove the affected microchip.
• Rinse the microchips according to the procedure described in "Inspecting and Washing Microchip Reservoirs".
  Click here for "(MCE) Inspecting and Washing Microchip Reservoirs (manually)".
• Perform maintenance on the chip cover.
  Click here for "(MCE) Maintenance of Chip Cover".

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.