(MCE) (ERROR)Chip # has an error

(MCE) (ERROR)Chip # has an error
Category : 

Answer

<Details>
The test current value for the chip that generated "fill check NG" was also outside the standard range after the separation buffer refill.
 
<Instrument Operations>
The microchip that generated the error is washed automatically and becomes unavailable in the software.
The analysis schedule is rescheduled and continues to execute only using the available microchips.
 
<Corrective Measures>
• After analysis, open the top cover and chip cover and remove the affected microchip.
• Confirm that it is unobstructed according to the procedure described in "Inspecting and Washing Microchip Reservoirs".
  Click here for "(MCE) Inspecting and Washing Microchip Reservoirs (manually)".
• To reuse a microchip that was unobstructed (or for which the blockage was resolved), clear the unavailable state according to the procedure described in "Chip Management" by instruction manual.
• Perform maintenance on the chip cover.
  Click here for "(MCE) Maintenance of Chip Cover".